1.BFG CODE NO.- BFG Code No.consists of minimum 1 and maximum 25 alpha-numeric and special characters.
2. PRODUCT NAME- Product Name consists of minimum 1 and maximum 250 alpha-numeric and special characters.
3. PRODUCT VERSION NO.- Product Version No. consists of minimum 1 and maximum 25 alpha-numeric and special characters.
4. VERSION NO.- Version No. consists of minimum 1 and maximum 25 alpha-numeric and special characters.
5. PRODUCT TYPE- Type of product to be checked from the options.
- CAPSULE FILLING: The process represents performing weighment of capsules considering different parameters.
- GRANULATION: The process of forming grains or granules from a powdery or solid substance, producing a granular material. In simple words, granulation involves shredding or grinding solid material into finer granules or pellets.
6. SET UNIT- Unit of Measurement to be selected for Balance and Vernier related parameters.
Click on 'Set Unit' button and Unit of Measurement window will get displayed allowing to select unit for Individual, Group and Dimension parameters.
Select required units for respective parameters and click on 'Save' button; the following window will get displayed notifying "Unit Saved Successfully".
- STD. VALUE: It is a nominal value of product for respective test.
- TOLERANCES: Negative and Positive tolerances of product for respective test.
- LIMIT ON: To be selected from dropdown list; can either be Actual (which means tolerances to be considered at actual value) or Percentage (which means tolerances to be considered as percentage of standard value).
- REPORT ON: To be selected from dropdown list; can either be Standard (which means graphical representation of test will be at/on standard value) or Average (wherein graphical representation of test will be on average value).
- NMT TOLERANCE 1: NMT or Not More Than Tolerance 1 indicates acceptable number of samples crossing ‘tolerance 1’ limits.
- NOMINAL NOMENCLATURE: Nomenclature to be selected from dropdown list.
- MIN/MAX TEMP.: Minimum/maximum temperature to be set and controlled; hereby monitor time within which capsule gets dissolved.
*NOTES:
- At least one parameter must be selected while adding the product so that its structural integrity can be monitored.
- Same product details cannot be added more than once; any one field must differ from existing details.
Put in all the required information and click on Save option; the following window will get displayed asking for confirmation.
Click on Cancel option and the Add Capsule page will get displayed with new filled information;
Click on Confirm option and the following window will get displayed asking to enter remark;
Enter remark and click on OK option; the following window will get displayed notifying Product Added Successfully;